©2017 Abbott Point of Care Inc. All rights reserved.Printed in USA. i-STAT is a registered trademark of theAbbott Group of Companies in various jurisdictions.i-STAT®Art: 714254-01S Rev. Date: 13-Feb-17Emergo EuropeMolenstraat 152513 BH The HagueThe NetherlandsAbbott Point of Care Inc.100 and 200 Abbott Park RoadAbbott Park, IL 60064 • USAFor information related to Article 33of the EU REACH regulation(EC No. 1907/2006), please referto PMIS.abbott.com. If you haveissues logging into the website,contact Abbott at: .......................abbott.REACH@abbott.com.Message Cause ActionAnalyzer Error, UseElectronicSimulatorHandheld detectsproblem from which itis likely to recover.Insert the Electronic Simulator. IfPASS, continue to use handheld.Analyzer Error,See ManualHandheld detectsproblem from which itmay not recover.Insert Electronic Simulator. If PASS,insert a cartridge with sampleor control. If the code does notreappear, continue to use handheld.No display Disposable batteriesdead or rechargeablebattery fullydischarged. Keypadnot responding. Startswitch broken.Change disposable batteries orrecharge battery. If still no display,call Support Services.Cartridge Lockeddoes not disappearafter test cyclecompletedDead battery(s).Mechanicalproblem.Wait until handheld turns off. Turnhandheld on. If resets, removecartridge. If not, change or rechargebattery(s) and turn handheld on.Quality Check Messages and Codes (continued)06F16-57(01)00054749002139i-STAT®1User Guide