Keysight Y1225A Rack Mount KitInstallationThis installation note contains procedures for installing the Y1225A Rack MountKit for the Keysight M9502A 2-slot AXIe Chassis in a standard EIA rack.Parts ListThe parts included with each rack mount adapter option and the support rail areshown in the following table:General GuidelinesIn rack mounting the chassis, follow these guidelines:1 Always begin installing the chassis and heavy instruments at the bottom ofthe rack and work up. This maintains a lower center of gravity and reduces thepossibility of the rack tipping.2 Anti-tipping feet, if available with the rack, should always be extended.3 The heaviest instrument of chassis should always be mounted at the bottomof the rack.4 For maximum cooling and optimum rack thermal efficiency, place theinstrument or chassis with the greatest power consumption towards the top ofthe rack. This promotes efficient cooling since heat rises when placed nearerto the top of the rack, high power instruments will not unnecessarily heatother instruments. However, in doing this, do not violate the guideline thatthe heaviest instruments should be placed at the bottom of the rack.5 If instruments or modules in the chassis are consuming maximum power, 1Uof rack space is required for ventilation below the chassis.Quantity Description Part Number1 Rack Flange Kit N5650-000204 Rack Flange Screws M4x6 Flat Head N/A2 Support Rails M9018-000018 Channel Nuts (for support rail-to-rack) 0590-08044 Support Rail-to-Rack Screws - 10-32 x 0.5 2680-02784 Dress Adapter Screws for rack flanges E9900-06001All manuals and user guides at all-guides.comall-guides.com