n ®QUICK START GUIDETESTING1. Press and hold until the meter turns on.2. Tap . Select a Water Source type. Tap to confirm.3. Tap . Select a disk series (found on disk packaging). Tapto confirm.NOTE: Disk Series are limited by Water Source Type selection.4. Remove a disk from the packaging.5. Use the syringe (1189) to fill the disk with the water sample.6. Insert the disk. Cover the disk with the Universal Disk Cover.Close the lid.7. Tap to start test. Tap to stop the test. If the test iscancelled discard the disk.8. The results will be displayed.9. Choose an option.¬ Tap to add a tag.¬ Tap the highlighted to save the test results to thetest log if Auto Save is not enabled.¬ Tap the hightlighted to send the results to theenabled Mobile Bluetooth Printer.¬ Tap the highlighted to send the results to aBluetooth enabled device.¬ Tap to return to the Test Screen.10. Press and hold for 2 seconds to turn the meter off.DOS & DON’TSBEFORE FIRST USE¬ Fully charge battery using the Spin TouchUSB cable and adapter plug in AC outlet.¬ Ensure your Spin Touch has latestfirmware1. Download and install WaterLink®Connect 2 Application for Windows athttp://softwarecenter.lamotte.com/.Select WaterLink® Connect 2 Softwarefor Windows FREE Download.2. Plug meter into the computer withthe provided USB cable and launchWaterLink Connect 2 Application fromthe Start Menu.3. A prompt will appear if firmwareupdates are available. Select Update.Testing and data transfer will notbe possible until firmware has beenupdated.NOTE: For water samples over 100 °F (38 °C) subtract0.3 from pH result or, for the most accurate result,wait until water sample is below 90 °F (32 °C) to test.Remove residue daily. Residue will damage the meterand cause inaccurate results. See Cleaning in manual.¬ Do not touch top or bottom of disk. Handle disk by theedge.¬ Do not fill disk while in the meter. Fill disk on clean, drysurface.¬ Fill the disk on a dark surface to more easily see thesample water.¬ The disk should not contain any large air bubbles. Airbubbles will result in erroneous results.¬ Always use the disk cover.¬ Only the Universal Disk Cover (Code 1719) can be usedwith the WaterLink® Spin Touch®.¬ Empty syringe of old sample before filling with nextsample.¬ Remove filled disk from meter after testing. Do nottravel with filled disks in meter. They may leak.¬ Keep the chamber clean and dry. Gently swab LEDand photodiode lenses located around the hub witha cotton swab dampened with streak-free windowcleaner. Do not use alcohol. It will leave a thin film overthe lenses when dry.¬ Go to SETTINGS>Other Settings to select hardness andnitrogen units. Nitrogen results can be reported in ppmNitrate-N units or ppm Nitrate units. 1 Nitrate-N unit =4.4 Nitrate units.