The EASYSPRAY device will continue to emit gas for a brief period after the thumb is removed from the clip/plunger.This delay helps to avoid clogging of the spray head.Please see the ARTISS SmPC provided together with this material.Instructions for Circulating Nurse | EasySpray Pressure RegulatorInstructions for Scrub Nurse | Spray SetPass the assembled applicator to the sur-geon for spray application. Pass the endof the connection tube with the sterilefilters to the circulating nurse.6Attach the clip (on the end of the sensorline) by sliding it into the grooves locatedon the top of the syringe plunger.5Fit the connection tube of the spray set tothe luer-lock connector on the underside ofthe spray head.4Fasten the pull strap to the doublesyringe system to assure the spray headis tightly secured.3Firmly attach the spray head to thenozzle of the syringes.2Prepare ARTISS Solution for Sealantsaccording to the instructions in thepackage insert.1Instructions for SurgeonTo activate the flow of gas occlude theopening in the clip center with thumb.To begin application, gently depress thesyringe plunger.3Spray from a distance of 10 – 15 cm foroptimum results.2Confirm (verbally) the actual pressurewith theatre personnel.1Insert 9V battery into the EASYSPRAYpressure regulator device.Check the gauge on the EASYSPRAYdevice for the appropriate pressurerange of 1.5-2.0 bars (21.5-28.5 psi).Adjust pressure setting by turning theblack pressure control knob.6Turn the on/off switch on the front sideof the EASYSPRAY to the ON (I) position.5Connect Spray Set filters to EASYSPRAYdevice. Connect the blue filter to the bluefemale luer connector and the clear filter tothe male luer connector.4Use suitable connection tube to connectthe EASYSPRAY device to medical air(ranging 3.5 – 7 bar / 50 – 100 psi).3Connect EASYSPRAY device to IV pole ortrolley rail using the clamps on the backof the device.21The use of ARTISS Solutions for Sealant is restricted to experienced surgeons who have been trained in the use of ARTISS.In order to ensure optimal safe use of ARTISS by spray application, apply a minimum spray distance of 10 cm and a maximum spraypressure of 2 bar to minimise the potential risk of air or gas embolism, tissue rupture, or air or gas entrapment with compression.easysPRay Quick Reference Guide (for open surgery)Caution: Any application of pressurised gas may be associated with a potential risk of air or gas embolism, tissue rupture or air or gasentrapment with compression, which may be life threatening. Be sure to take appropriate measures to address these risks by observingthe recommended minimum spraying distance and the maximum pressure provided in the appropriate spray set instructions for use.