Document Number: D1000092206 B 152API 5000™ Instrument ParametersThe following table contains generic parameters for the API 5000™ instrument.The first number under each scan type is the preset value; the range of numbersis the accessible range for each parameter.Table 2-1 API 5000™ Instrument ParametersParameterIDAccessIDPositive ion mode Negative ion modeQ1 Q3 MS/MS Q1 Q3 MS/MSCUR CUR 10 10 10 10 10 1010 to 50 10 to 50 10 to 50 10 to 50 10 to 50 10 to 50CAD CAD 0 1 4 0 1 4n/a 0 to 12 0 to 10 n/a 0 to 12 0 to 12IS (1)(2) IS (1)(2) 5500 5500 5500 –4500 –4500 –45000 to 5500 0 to 5500 0 to5500–4500 to0–4500 to0–4500 to0NC (3) NC (3) 3 3 3 –3 –3 –30 to 5 0 to 5 0 to 5 –5 to 0 –5 to 0 –5 to 0TEM (2)(3) TEM (2)(3) 0 0 0 0 0 00 to 750 0 to 750 0 to 750 0 to 750 0 to 750 0 to 750OR DP 120 120 120 –100 –100 –100(DP = OR) 0 to 400 0 to 400 0 to 400 –400 to0–400 to0–400 to 0Q0 EP 10 10 10 –10 –10 –10(EP = –Q0) 15 to 2 15 to 2 15 to 2 –15 to–2–15 to–2–15 to –2IQ1 IQ1 Q0 + (–1) Q0 + (–1) Q0 +(–1)Q0 + 1 Q0 + 1 Q0+1(IQ1 = Q0 + offset) –0.5 to –2 –0.5 to –2 –0.5 to–20.5 to 2 0.5 to 2 0.5 to 2ST ST Q0 + (–7) Q0 + (–7) Q0 +(–7)Q0 + 7 Q0 + 7 Q0 + 7(ST = Q0 + offset) –8 to –4 –8 to –4 –8 to –4 4 to 8 4 to 8 4 to 8RO1 IE1 1 n/a 1 –1 n/a –1(IE1 = Q0 – RO1) 0.5 to 2 0.5 to 2 –2 to–0.5–2 to–0.5RO1 RO1 n/a Q0 + (–2) n/a n/a Q0 + 2 n/a(IE1 = Q0 + offset) –0.5 to –2 0.5 to 2(1)IonSpray™ ion source (2)TurboIonSpray® ion source (3)Heated Nebulizer (4) 1=ONand 0=OFF (5)DuoSpray™ ion source, 1=TIS, and 2=HN