26Alarm 1 FailsafeSelect Yes to enable failsafe action, otherwise select No.See also Figs. 5.1 to 5.5 (page 30).Alarm 1 ActionSelect the alarm action required, High or Low.See also Figs. 5.1 to 5.5 (page 30).Alarm 1 Set PointSet the alarm set point to the required value within the sensor's measurement range – seeSensor Group on page 17.Initial ChargeSelect Yes to enable an initial charge sequence. The initial charge sequence commenceswhen A: Initial Charge is set manually to Yes in the Operating Page (Section 2.3.1) andcontinues for the initial charge time period set below.Initial Charge TimeSet the initial charge period, between 1 and 30 minutes in 1 minute increments.Continued on next page.A1: ActionLowHigh-----A1: FailsafeNoYes-----Initial ChargeInit. Charge Time30MinsA1: Setpoint1000mS/cmA1: Typeset to DosingNoYesDosingYesNo…5.4 Configure Alarms5.4.2 Configure CIP Interface AlarmNotes.• This section is applicable only if A1: Type is set to Dosing – see Section 5.4.1.• Alarm 4 (dual input analyzers only with option board fitted and analog features enabled – see Section 7.3) can beconfigured as a CIP Interface alarm for Sensor B, therefore this Section applies also to Alarm 4.• When Alarm 1 (and/or Alarm 4) is assigned as a CIP Interface alarm, the error messages shown in Table 5.1 on page 29are displayed in response to the events described.…5 PROGRAMMING