LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A 79Observing cable specificationsThe connection cables are components of the transmitter unit and receiverunit. They are preinstalled at the factory.Note: The connection cables must only be replaced with genuine replacementcables.Cable type 8 x 2 x 0.5 mm 2 ; weather-resistantOuter diameter 6.5 to 14 mmOuter diameter Min. 7 mm up to max. 12 mmWire cross-section Min. 0.5 mm2 to max. 2.5 mm 2 (AWG 20–14)Version With shieldOuter diameter Min. 7 mm up to max. 12 mmSuitable for cable gland M20Wire cross-section 3 x min. 1.5 mm2 to max. 2.5 mm2 (AWG 16–14)Laying of single wires is only permitted in theconduitFlammability class VW-1 and FT-1Temperature resistance Ambient temperature > 60 °CCable gland M20Version Fitting with integrated cable seal and option toconnect a conduitConduit outside diameter 19 mmCable cross-section Max. 4 mm 2 ; typ. 2.5 mm 2(AWG 8 in accordance with CSA regulations)Cable length As short as possible (minimum length 1 m)Connection cablesfor the transmitterunit and receiver unitConnection cablesfor T/P probesConnection cablesfor the power supplyCable gland for thepower supplyPotentialequalization cablesfor transmitter unit,receiver unit andjunction box