30 TTF200 FIELD-MOUNT TEMPERATURE TRANSMITTER | OI/TTF200-EN REV. A9.1 Possible HART error messagesRange DisplaysthedevicestatusDisplaysthe DIAG.NO.Cause RemedyElectronics F 1 Device defective Replace the deviceElectronics S 2 Ambient temperature overshot /undershotCheck environment; reposition measuringpoint if requiredElectronics F 3 EEPROM defective Replace the deviceElectronics M 4 Electronics overload Factory resetElectronics F 5 Memory error Factory resetElectronics I 7 LCD display connected Status info; not an errorInstallation /ConfigurationI 8 Device write-protected Status info; not an errorElectronics I 9 EEPROM busy Status info; not an errorElectronics F 12 Sensor input defective (communication) Replace the deviceElectronics F 13 Sensor input defective (error) Replace the deviceElectronics F 14 Sensor input defective (ADC error) Replace the deviceInstallation /ConfigurationC 32 Diagnostics simulation mode Not an error, diagnostic info, measurementOKSensor F 34 Measuring error, sensor Check sensor connectionSensor F 35 Short-circuit, sensor Check sensor connectionSensor F 36 Wire break, sensor Check sensor connectionSensor F 37 Range overshot, sensor Check measuring limitsSensor F 38 Range undershot, sensor Check measuring limitsInstallation /ConfigurationI 41 Single-point adjustment active, sensor Status info; not an errorInstallation /ConfigurationI 42 Two-point adjustment active, sensor Status info; not an errorInstallation /ConfigurationF 65 Configuration defective Check configuration:A) Wrong deviceB) Span is too smallC) Incorrect configuration dataInstallation /ConfigurationC 71 Reconfiguration is running Status info; not an errorOperating conditions F 72 Error in the application Check configuration, connections; reset tofactory settingsInstallation /ConfigurationI 74 Analog output adjustment active Status info; not an errorInstallation /ConfigurationC 75 Analog output in simulation Status info; not an errorOperating conditions S 76 Values overshot Check parameters:A) Sensor limits overshotB) Span is too smallNOTICEIf the remedial measures listed for the error message do not improve the status of the device, please consult ABB Service.Change from one to two columns