The instrument is to be cleaned only by wiping Virkon solution at the manufacturer’s guidelineconcentration. No parts are to be autoclaved.96. Cleaning >>7. Specification >>Equipment Reference: ALPS 50V™Model No: 953-ASpecification:Temperature Set Point Range 125°C to 200°C in 1°C increments.Time Set Range 1 to 9 seconds in 0.5 second increments.Operating Humidity Range 20% to 80% non-condensing.Operating Temperature Range 18°C to 30°C.Size 220 x 321 x 425 mmWeight 7.2 kgPower Supply 100 - 130V AC 50/60Hz or 220 – 240V AC 50/60HzPower Rating 350 W MaxFuse 100 - 130V - T3.15A ( IEC 127 ) or220 - 240V - T1.6A (IEC 127)Mains Input Connector IEC 320Applicable Standards FCC Class A, CFR 47 Part 15BS EN 61326:19988. Over Temperature Safety Cut Out >>The instrument is protected against a fault condition leading to the sealing plate heating upto a temperature exceeding it’s maximum set point. Protection is provided by a single usethermal cut out, operating in the range 235°C to 240°C. In the unlikely event that this cut outshould operate, it would necessitate the sealer being returned to the manufacturer for repair.