DC/AC MicroProbe Models K100 and K110 215.3 CleaningTo ensure optimum performance, it is important to keep the probe jawmating surfaces clean at all times. Failure to do so may result in increasedearth’s field susceptibility and overall errors in readings. The following isthe recommended procedure for cleaning the probe sensor:(1) Apply a few drops of isopropyl alcohol to an ordinary sheet of whitephotocopy paper.(2) Open probe to expose the mating surfaces of the sensor and insertthe sheet of paper into the jaws. Allow the jaws to close.(3) With the jaws closed, pull free the end of the paper through the jawsuntil it releases.(4) Repeat step (3) several times, using a new (dry) section of papereach time.(5) Open probe and inspect mating surfaces. They should be free ofdust particles or any other contamination. Otherwise, repeat thecleaning process starting again at step (1).ALCOHOL1234Figure 17www.ShopAEMC.comShop for AEMC products online at: 1.877.766.5412