19UnitAD-4212B-102 AD-4212B-201 AD-4212B-301Capacity MinimumdisplayCapacity MinimumdisplayCapacity MinimumdisplayGram 110 0.00001 210 0.0001 310 0.0001Milligram 110000 0.01 210000 0.1 310000 0.1Ounce (Avoir) 3.88 0.000001 7.40 0.00001 10.93 0.00001Troy Ounce 3.53 0.000001 6.75 0.00001 9.96 0.00001Metric Carat 550 0.0001 1050 0.001 1550 0.001Momme 29.3 0.00001 56.0 0.0001 82.6 0.0001Pennyweight 70.7 0.00001 135.0 0.0001 199.3 0.0001Grain (UK) 1697 0.0002 3240 0.002 4784 0.002Tael(HK general, Singapore) 2.91 0.000001 5.55 0.00001 8.20 0.00001Tael (HK jewelry) 2.93 0.000001 5.61 0.00001 8.28 0.00001Tael (Taiwan) 2.93 0.000001 5.60 0.00001 8.26 0.00001Tael (China) 3.52 0.000001 6.72 0.00001 9.92 0.00001Tola (India) 9.43 0.000001 18.00 0.00001 26.57 0.00001Messghal 23.4 0.00001 44.8 0.0001 66.1 0.00015-2 Changing the UnitA unit or mode can be selected and stored in the function table.Select a unit or mode as follows:1 Press and hold the SAMPLE key until ba5fnc of thefunction table is displayed, then release the key.2 Press the SAMPLE key several times to displayUnit .3 Press the PRINT key to enter the unit selection mode.4 Press the SAMPLE key to select a unit or mode to beused for weighing.5 Press the RE-ZERO key to specify the unit or modeselected in step 4.(The stabilization indicator illuminates.)6 Press the PRINT key to store the unit or mode. Thebalance displays end and then displays the nextmenu item of the function table.7 Press the CAL key to exit the function table. Then thebalance returns to the weighing mode with the selectedunit or mode.SelectSpecifyStoreDisplays the unitssequentially.