May 2014 – TD 293 OPERATING MANUAL – SmartGuard Page 496.5 AADI SR10/VR22 sensorsOpen SR10/VR22 Sensors in the DeviceLayout.Press Add below the list of sensors, referFigure 6-15, and enter sensorinformation.AADI SR10/VR22 sensors require the5120 or 5320 versions of SmartGuardwhich are equipped with electronicinterface for these sensors.There is one connector capable ofconnecting up to 11 SR10/VR22 sensorsthrough a variety of different cablesdepending on the actual application,refer page Error! Bookmark notdefined..The signals are connected to Channel 1through 11 by the cable layout and the user must keep track of which parameter appears on which channel.All actual SR10 and VR22 sensors are preconfigured in SmartGuard. The user only needs to tell whichchannel to read and the actual sensors serial number.All SR10/VR22 sensors are power controlled as one unit which implies that those sensors cannot bedistributed over multiple recording groups; they must all be included in the same recording group.6.5.1 AADI SR10/VR22 Sensor layoutFigure 6-16. Press Edit to change existing layout.Figure 6-15 SR10/VR22 Sensors.Press Add below the list of SR10/VR22 sensors, and enter sensor information as shown in