CELL-DYN® 1800 OPTICON OPT-6125 Trigger-Activated Hand-Held Bar Code Scanner User’s Guide9140394A—November 2008 1Scanner Installation:Materials Required:• OPTICON OPT-6125 Bar Code Scanner• Adapter Cable, 5-Pin AT to 6-Pin PS/2• Bar Code Scanner Holder (optional)1. From the MAIN MENU press SPECIAL PROTOCOLS, then DAILYSHUTDOWN. Wait for the sequence to complete (approximately 5minutes).2. Power OFF the CELL-DYN 1800 System.A. Install New Bar Code Scanner1. Disconnect the keyboard cable from the right side of theCELL-DYN 1800 System.OPTICON OPT-6125 Barcode Scanner with Adapter Cable2. Plug the CELL-DYN 1800 Keyboard Cable into the Scanner Cable at the5-pin AT connector, making sure the pins line up.3. Plug the 6-pin PS/2 connector on the Adapter Cable into the matchingconnector on the Scanner Cable, making sure the pins line up.4. Plug the 5-pin AT connector on the Adapter Cable into the connector on theright side of the CELL-DYN 1800 System, making sure the pins line up.5. Attach the (optional) Bar Code Scanner Holder to the right side of theCELL-DYN 1800 System by removing the adhesive covering andpressing the hanger firmly to the instrument.6. Power ON the CELL-DYN 1800 System, wait for the INITIALIZEDmessage to appear in the Status Box, then press PRIME/RUN to bring theCELL-DYN to READY.1. HandheldBarcodeScanner2. Adapter Cable3. 6-Pin PS/2connector4. 5-Pin ATconnector123 4