19Energy Factors for Tonic Stimulation ParametersThe following tables show energy factors according to various stimulation parameters for tonicprograms.NOTE: Energy factors are for IPGs that provide 12 hours of daily stimulation. For an IPG thatis providing 24 hours of daily stimulation, double the energy factor shown in the table.Table 6. Energy factors for various tonic stimulation parameters (350-ohm impedance)Pulse Width (μs)Amplitude (mA) Frequency (Hz) 100 200 300 50030 13 14 14 151 60 18 19 20 2390 22 24 26 2930 15 17 20 242 60 21 26 30 4090 27 34 41 5430 16 19 23 303 60 24 30 37 5190 30 41 51 7130 17 22 26 474 60 26 35 44 8590 34 47 61 12230 21 30 38 555 60 34 51 68 10290 46 71 97 14830 23 33 43 816 60 37 58 78 15290 51 81 112 22330 29 44 60 927 60 49 80 112 17590 68 115 162 25730 31 49 67 1258 60 53 89 125 24290 74 129 183 35930 33 64 89 1399 60 58 118 169 27190 81 173 249 40130 41 69 97 18210 60 73 130 186 35590 105 190 274 528