2824. Electromagnetic conditionsSystem specification:- D-1090-2G monitor- L-4060SL or L-4060SLC bed sensor- GlobTek external power supply (see chapter Technical Specifications for details)Cable specification:- Power cable (non-shielded) max. Length 2 m- Sensor cable (shielded) max. length 3 mNote! RF communications equipment can effect medical electrical equipment!Guidance and manufacturer’s declaration – electromagnetic emissionsThe Emfit Tonic-Clonic Seizure Monitor is intended for use in the electromagneticenvironment specified below. The customer or the user of the equipment shouldassure that it is used in such an environment.RF emissions CISPR 11 Group 1The Emfit Tonic-Clonic Seizure Monitor usesRF energy only for its internal function.Therefore, its RF emissions are very low andare not likely to cause any interference innearby electronic equipment.RF emissions CISPR 11 Class B The Emfit Tonic-Clonic Seizure Monitor issuitable for use in all establishments,including domestic establishments and thosedirectly connected to the public low-voltagepower supply network that supplies buildingsused for domestic purposes.Harmonic emissions IEC61000-3-2 Class AVoltage fluctuations/flicker emissions IEC61000-3-3Complies