9PRODUCT INFORMATIONCorrect Usage Operation of the A C C E N T may only be carried out by suitablyqualified personnel. The A C C E N T is only to be used in the production of intraoralX-ray exposures. The A C C E N T should be used in a room equipped for it. Room temperature should be in the range 50 to 105°F (10 to 40°C)with relative humidity between 20 and 60%. If the device is brought into the room of operation from a coolerenvironment, condensation can build up. Do not connect thedevice until it has warmed up to room temperature and isabsolutely dry. This room should be free of all possible interferences (e.g. strongmagnetic fields), as these could affect the operation. The A C C E N T may only be operated together with authorizedsoftware. Correct usage includes observing all installation and operatinginstructions and adherence to the set-up, operation and main-tenance instructions. Any use, above and beyond that described in this manual ascorrect usage, will invalidate the warranty. The operator is liable for all damages caused as a result ofincorrect usage of this appliance. Only A C C E N T Sensors made by Air Techniques may be usedwith the A C C E N T Control Module.Incorrect Usage Any use that is not described in this manual as correct usage isconsidered as incorrect usage. The manufacturer is not to be heldliable for any damage caused as a result of incorrect usage. Theoperator bears all risks in this case. Additionally, the manufacturer is not liable for any damage tothe A C C E N T Sensor caused by patient biting, dropping orincorrect handling. See warranty on page 3 for the A C C E N TSensor.