Using the meter as a diary155If the transfer is unsuccessfulIf the results are not transferred, the meter turns offautomatically after about 60 seconds.Possible reasons for an unsuccessful transfer are:A transfer error occurred.Start the transfer of the results again.The infrared windows are too far apart or not pro-perly facing each other.Reduce the distance between the two infraredwindows to 5–20 cm and position them so they arepointing towards one another. Start the transfer ofthe results again.Another device with an infrared source turned on, ora strong light source is too close to one of the twoinfrared windows.Remove the device or light source and start thetransfer of the results again.There is some object, for example, blocking commu-nication between the two infrared windows.Remove the object and start the transfer of theresults again.An error occurred at the receiving end.Please read the relevant instructions for use.