6.4 Connecting and disconnecting M+ accessories during useDo not connect/disconnect the cords when the medical device is switched ON and the footswitch ispressed.Do not tighten or loosen the contra-angles when the I-Surge LED micromotor is activated.Do not tighten or loosen the tips when the handpiece is activated.6.5 Switching off the medical deviceTo avoid damaging the I-Surge LED micromotor, it must be rinsed with sterile water less than 30minutes after use.1. Remove the bag from the bracket.2. Remove the irrigation line perforator from the bag.3. Immerse the irrigation line perforator in a container with distilled water.4. Press and run the purge cycle for one minute to rinse the handpiece.5. Remove the perforator from the container and purge the system until the medical device's irrigation lineis completely empty.6. Switch off the medical device (O/I switch to O).User Manual • ImplantCenter™ M+ • J27251 • V1 • (13) • 09/2013 • NO25EN010A - Page 19/42