6actigraphcorp.comThis section describes how to configure the wGT3X-BT and begin collecting data.Open the ActiLife software.Connect the wGT3X-BT to the PC using the USB cable. The monitor(s) will appear in the grid under the ‘Devices’ tab.Select ‘Initialize’. A submenu will open to display several initialization options.Select ‘Regular Initialization’ from the submenu. A dialog box will open to display the initialization parameterslisted below. The most commonly used initialization parameters are pre-selected as system defaults.Initialization ParametersStart and Stop TimesEnter the dates and times when the device should start and stop collecting data. If no start time is selected, thesystem defaults to two minutes ahead of the current time. If no stop time is selected, the device will continueto collect data until the battery is depleted, the memory capacity is full, or it is downloaded and reinitialized.1234Setup & InitializationMoses (1)