18The SA-2010S provides ECG Waveforms to aid in the verification of the basic operationalcharacteristics of cardiac-monitoring equipment. The information in this section presents alisting of available waveforms. It is the responsibility of the user to establish whichwaveform verifications are required based on local codes, facilities practices and equipmentmanufacturer’s recommendations.Pulse WavesThe SA-2010S simulates a Normal Sinus Rhythm pulse at 30, 60, 120 and 240 BPM.With ECG Waveforms function selected, use the arrow keys under the waveform LEDs toselect the desired pulse. The display is unused in this mode and will show over range. TheLEDs will pulse with the waveform.Performance WavesSineThe SA-2010S sends out Sine waves of 10, 60 and 100 Hz. With ECG Waveforms functionselected, use the arrow keys under the waveform LEDs to select the desired Sine wave.The display is unused in this mode and will show over range.SquareThe SA-2010S sends out Square waves of 0.125 and 2 Hz. With ECG Waveforms functionselected, use the arrow keys under the waveform LEDs to select the desired Square wave.The display is unused in this mode and will show over range.TriangleThe SA-2010S sends out Triangle waves of 2 Hz. With ECG Waveforms function selected,use the arrow keys under the waveform LEDs to select the desired Triangle wave. Thedisplay is unused in this mode and will show over range.ECG WAVEFORMS