15The SA-2010S Models provide ECG and Performance Waveforms to aid in theverification of the basic operational characteristics of cardiac-monitoring equipment.The information in this section presents a listing of available waveforms. It is theresponsibility of the user to establish which waveform verifications are required.Note: The LCD display is unused in these modes and will indicate an over-rangecondition shown as “1_ _ _”.Pulse WavesThe SA-2010S simulates a Normal Sinus Rhythm pulse at 30, 60, 120 and 240 BPM.With ECG Waveforms function selected, use the arrow keys under the waveform LEDsto select the desired pulse. The LEDs will pulse with the waveform.Performance WavesSineThe SA-2010S sends out Sine waves of selectable frequency - 10, 60 and 100 Hz.With ECG Waveforms function selected, use the arrow keys under the waveform LEDsto select the desired Sine wave output mode.SquareThe SA-2010S sends out Square waves of selectable frequency - 0.125 and 2 Hz. WithECG Waveforms function selected, use the arrow keys under the waveform LEDs toselect the desired Square wave output mode.TriangleThe SA-2010S sends out Triangle waves of frequency 2 Hz. With ECG Waveformsfunction selected, use the arrow keys under the waveform LEDs to select the Trianglewave output mode.PATIENT SIMULATORNOTICE – PERFORMING TESTSREFER TO DUT MANUFACTURER’S SERVICE MANUAL FORTEST PROCEDURES AND MEASUREMENT LIMITS.