17Technical ManualWarning: The use of accessories other than those specified in Section 2.3 of the Site~Rite Vision* Ultrasound System Hardwareand Operating System instructions for use may result in increased emissions or decreased immunity of the Site~RiteVision* Ultrasound System.Note: While the Site~Rite Vision* Ultrasound System is not classified to be used in a domestic environment, there may be a needfor the health care professional to use the system for a limited duration.Warning: This system, with its applicable accessories, is intended for use by healthcare professionals only. If used in a domesticenvironment, this system may cause radio interference or may disrupt the operation of nearby equipment. It may benecessary to take mitigation measures such as re-orienting or relocating the Site~Rite Vision* Ultrasound System orshielding the location.Technical SpecificationsRF Emissions CISPR 11RF Emissions CISPR 11Harmonics IEC 61000-3-2Flicker IEC 61000-3-3Group 1Class AClass ACompliesThe Site~Rite Vision* Ultrasound System uses RFenergy only for its internal function. Therefore, its RFemissions are very low and are not likely to cause anyinterference in nearby electronic equipment.The Site~Rite Vision* Ultrasound System is suitablefor use in all establishments, other than domestic,and those directly connected to the public low-voltagepower supply network that supplies buildings used fordomestic purposes.Electromagnetic Environment – GuidanceEmissions Test ComplianceEMC Tables4.2The Site~Rite Vision* Ultrasound System is intended for use in the electromagnetic environment specified below.The customer or user of the Site~Rite Vision* Ultrasound System should ensure that it is used in such an environment.GUIDANCE AND MANUFACTURER’S DECLARATION - EMISSIONSSite~Rite Vision* Ultrasound System EMC Tables