34SPECIFICATIONS (Continued)Thermistors:YSI 400: −40 to 100°C ( −40 to 212°F)YSI 700: −30 to 100°C ( −22 to 212°F)Platinum RTD100 ohm: −200 to 850°C (−328 to 1562°F)INPUTSAC Line Input115V Model 230V ModelInput: 115V AC ±15%, 49/61 Hz 230V AC ±15%, 49/61 Hz15A Max 15A MaxConnector: Standard Line Cord IEC 320 C-20 InletFuse: 0.630A/230V Slo-Blo 0.315A/230V Slo-Blo(T) 5x20mm (T) 5x20 mmInstallation Category: Category II per IEC664(Local level — appliances, portable equipment, etc.)Thermocouple Probe InputConnector: Mini-ANSI flat blade thermocouple jackCalibration: ITS-90 (NIST monograph 175)Lead resistance: 500 ohm maxCommon mode voltage: 5 V maxThermistor Probe InputAny of the YSI series 400 and 700 series probes and YSI 44004, 44033,44018 and 44104 composite thermistor elements.Connector: 3-pin 1/4 in phone jackCONTACT 400 series 700 seriesTIP: Therm. Therm. (T1)RING: not used Therm. (T2)BARREL: Therm. Therms. (Common)400 series 700 seriesCalibration: YSI 44033 YSI 44018Impedance: 2252 ohm @ 25°C (77°F) (T1) 6,000 ohm @ 25°C (77°F)(T2) 30,000 ohm @ 25°C (77°F)Interchangeability: ±0.10°C ±0.15°CLead Resistance: 0.45 ohm max 1.2 ohm maxPower Dissipation: 0.5 mW max 1.3 mW max