12 - ENG PN MN236 A | User ManualSprint / Stroller / Hi Flow StrollerCleaning StandardWARNING: CLEAN ONLY AFTER THE UNITIS EMPTY.• Clean using a solution of mild dish washingdetergent and water.• Apply cleaning solution directly to a lint-freecloth. Approved cleaners include HydroPureand HydroKlean. Do not spray cleaners di-rectly on the Sprint/Stroller/Hi Flow Stroller.• Wipe the outside surface with the lint-freecloth until the outside surface is clean.Caution: Do not use high temperatureand high pressure washing equipmentto clean these units.• Do not get cleaner on any internal compo-nents or valves.• Allow the unit to dry thoroughly before using.Note: Note to health care provider – for reprocessingprocedures, see applicable service manual.WEEE and RoHSThis symbol is to remind the equipmentowners to return it to a recycling facility atthe end of its life, per Waste Electrical andElectronic Equipment (WEEE) Directive.Our products will comply with the restriction ofHazardous Substances (RoHS) directive. They willnot contain more than trace amounts of lead orother hazardous material content.DisposalAlways return Sprint/Stroller/Hi Flow Stroller, in-cluding all components, to your homecare providerfor proper disposal. You can also contact yourlocal city or town offices for instructions on properdisposal of the battery.Transport and StorageThe device should be stored in the upright position,and be well ventilated. Do not allow the device tolie on its side. Humidity up to 95% noncondensing.Temperatures range from -40°F to 158°F (-40°C to70°C).Operating temperature ranges from 14°F to 104°F(-10°C to 40°C). Relative humidity range from30% to 75% noncondensing.Note: The atmospheric pressure range is 700 hPa to1060 hPa (elevation of 10,000 Ft. to -1,000 Ft.).MaintenanceYour service provider is responsible for any mainte-nance that my be required per the technical manualof this device. Call your service provider for anymaintenance requirements.The condensate pad is the only user-serviceablepart. No other parts are user-serviceable.The expected service life is a minimum of fiveyears.Clean the fill connectors on both the stationaryand portable units with a clean, dry, lint-free clothbetween each fill to prevent freezing and possibleequipment failure.Note: Any additional maintenance needed must bedone by a qualified service technician or serviceprovider.AccessoriesCarrying Cart PN 20765565