3CHAPTER TWOQUICK START GUIDEThis section will cover a basic quick start guide to initially get up and running with theDFX-7 for demonstration purposes only. However, in order to use the gauge andfeatures for actual applications, a thorough review of this manual is recommended.2.1 Gauge TypePower up the DFX-7, by pressing the on/off key located in the bottom right corner ofthe keypad. During initial boot up, a flash screen and lights will be displayed,followed by a GAUGE TYPE selection screen. Since this manual will focus on theflaw detector portion of the DFX-7, use the UP/DOWN arrow keys to select “flawdetector”, followed by pressing the OK key start the DFX-7 in flaw mode. If this is thefirst time the gauge has been used, a general startup configuration will automaticallybe loaded and the DFX-7 is ready for general inspections. If this is not the first timethe gauge has been used, and the user would like to reload the basic default setupparameters, refer to section 13.5 for further info.Note: The default setup should only be considered as a basic setup. Propercalibration and parameter settings should be done prior to an actual inspection.2.2 Basic Flaw Setup ReferenceThe following table lists all the settings for the basic flaw setup:PRB CAL DISP TUNE GT1Type: Single Matl: Steel 1020 View: +Rect Filters: Wide Alarm 1: On