SERIES 4 & SERIES 5 12 O’CLOCK CABINETInstallation InstructionsSITE PREPARATION5FIRST STEPSAIRPipe: 1/2” n.p.t. Protruding 1” from the floor. Supplied by contractor.Air supply: pressure 80-100 psi. Air should be free from contaminants.Pipe: 1/2” n.p.t. Protruding 1” from the floor. Supplied by contractor.Water supply: pressure 40-80 psi. Water should be flushed clean before connecting to this equipment.ELECTRICALElectrical: 4 x 4 x 2-1/4” Connection Box with 1/2” conduit knockouts.Wire box as per local code with top of the box no higher than 4-1/2” above finished floor.Voltage: 115 VAC, 3 wire, 15 or 20 amp with ground fault circuit interrupter (Series 5 cabinet only).Pipe: supplied by contractor. Terminated with 5/8 OD tube protruding 1” from the finished floor.Pipe: 5/8” OD tube protruding 1” from the floor. Place trap in-line and use vented fitting to conform to local codes.Supplied by contractor.CENTRAL VACUUMGRAVITY DRAINWATER1 Lay the installation template over plumbing stub-ups and check the position in the operatory.2 Install air and water shut-off valves (not supplied).3 Install 4 x 4 electrical connection box.4 Mark the cabinet mounting points according to the template.WARNING: Ensure the cabinet isproperly grounded to protective earth.WARNING: 3rd party x-ray systems must beconnected to a dedicated electrical circuit.CAUTION: Electrical utilities must be installed according to local codes and by a licensed electrician.See “Electrical Outlets” Page 11