Cerebral State Monitor - CSM DFU Page 104.7 SettingsParameter Menu Description Values (Default)High Alarm Alarm Selects the CSI Alarm High Option ON/OFF (OFF)Low Alarm Alarm Selects the CSI Alarm Low Option ON/OFF (OFF)High Limit Alarm Selects the CSI High Alarm Level 2-99(60)Low Limit Alarm Selects the CSI Low Alarm Level 1-98(40)Language Config Selects device language EnglishContrast Config Selects display contrast 15-35(20)RF interface Config Select radio link active. If turnedOFF the link is completely inactive.ON/OFF (ON)Set clock Config Sets clock: Date Format, Year,Month, Date, Hours, Minutes,SecondsBack light Config Sets display light and lightbrightness0 = Off1 = Normal (1)2 = BrightAutoLNK Config Automatically establishes awireless link to any CSM link withinapprox. 10 feet or to a specifiedCSM link.Inactive/Active/SN(Inactive) SN isonly displayed iflink is establishedThe AutoLNK functionWhen AutoLNK is set to Inactive:- active CSM Link(s) within range will pop-up in the Link display. Select the CSM Link serialnumber you would like to connect to.When AutoLNK is set to Active:- active CSM Link within range will automatically be connected to CSM monitorWhen AutoLNK is set to SN:- CSM monitor automatically connects to the dedicated CSM Link if within range.