Copyright © 2008 – 2019 DARE!! Products Page 205.2 Commands Plug-In CardThe following table shows the general commands for the RadiPower® plug-in card.Command Reply Description“*IDN?” “USB1004A, Version4.1.0”Returns the model and version ofthe device."ID_NUMBER?" “d.d.d.d.d.d.d.d” Returns the unique identifiernumber of the RadiPower®.“POWER?” “-12.34 dBm” Get measure power level in dBm.“BURST? ” “-12.34 -12.35 dBm”Get multi measure power level indBm.The number of measurements isrequested in ."FREQUENCY?" Hz Get the frequency in Hz."FREQUENCY Hz" “OK” or error code. Set the frequency in Hz."FREQUENCY? MIN" “9000 Hz” Get the minimal frequency."FREQUENCY? MAX" “6000000000 Hz” Get the maximum frequency."FILTER?" “1”, “2”, “3”, “4”, “5”,“6”, “7” or “AUTO” Get Filter."FILTER " “OK” Set Filter to 1,2,3,4,5,6 or 7."FILTER AUTO" “OK” Set Filter to auto.∑ IEEE command for Interface Clear or Clear are not supported.∑ IEEE Status flags in either serial or parallel poll, or as a service request, are not supported.∑ When IEEE communication is used, the first command/request should be the "*IDN?\r"command.