16A q u a - B o x O w n e r ’ s M a n u a lRIGID CANISTERS1. Transport full canisters by cart or roll-stand to the Aqua-Box using PPE’s in accordance with hospital policy.2. Ensure that the appropriate Evacuation Tube is attachedto the end of the suction hose (C).3. If fluid treatment option is used, verify that there issodium hypochlorite (bleach) and that the yellow tube (F)is submerged in the bleach. Check to see if the valve (E) forbleach line is open (vertically oriented).4. Open pour spout of the suction canister(s).5. Insert the Evacuation Tube and seal the port with thepurple cap.6. Turn the unit “On”.a. For Models 1500 and 1300, press the “Start” button(A). If the Aqua-Box shuts off while evacuating,simply press “Start” button (A) again.b. For Models 1200 and 1100, turn the valve on theright side of the unit to the “On” position.7. Open Thumb valve (D) to begin evacuation.INSTRUCTIONS FOR USENON-RINSING UNITSA - “Start” SwitchB - “Kill” SwitchC - Suction Hose (Grey)D - Thumb ValveE - Treatment Solution ValveF - Treatment Solution Tube (Yellow)G - Suction Hose HangerA BDGCEFWarning: As whenever handling any infections liquid waste, PPE’s should be worn at all times in accordance with facility policies.