Trigno™ Wireless Biofeedback System- 29 -Reference SpecificationsRF Frequency Band 2400-2483 MHz (ISM band)Sensor Dimension 27 x 37 x 13 mm (excluding leads)Mass 14 gTemperature Range(1) 5 - 45 degrees CelsiusTransmission Range 40 meters(2)EMG Signal Input Range 11mV r.t.i.EMG Signal Bandwidth 20-450 HzEMG Contact Dimensions 0.5 mmContact Material(3) 99.99% silverAccelerometer Range ±2g, ±4g, ±8g, ±16gAccelerometer Bandwidth 24 Hz – 470Hz (configurable in software)Gyroscope Range ±250 dps, ±500 dps. ±1000dps, ±2000dpsGyroscope Bandwidth 24Hz – 360 Hz (configurable in software)Inter-Sensor Delay < 1 sample period (Base Station only)Intra-Channel Delay < 1-2 sample periodAnalog Output Range ± 5 V (Base Station only)Analog Output Bandwidth (Ch. 1-16) DC-500 Hz (Base Station OnlyAnalog Output Bandwidth (Ch. 17-64) DC-50 Hz (Base Station Only)Analog Output Group Delay (Ch. 1-16) 48 ms (Base Station Only)Analog Output Group Delay (Ch. 17-64) 96 ms (Base Station Only)Software Compatibility EMGworks, Galileo Analytics, APIBattery Life 2 – 3 hours3rd-party Integration Support Analog outputs, Digital SDK/API support1) Exposure beyond these temperature limits may damage the rechargeable battery.2) Transmission range determined as maximum distance between two supported recording locations.3) Sensor skin contacts are made from pure silver and should not be used if allergic reactions to silver are expected or found to occur.