2INM MTL 130-0175 Rev 41.4 InformationWaste Electrical and Electronic Equipment directive (WEEE) 2002/96/EC(RoHS) directive 2002/95/ECWARNINGThis equipment must only be used in accordance with the manufacturer’sspecification, instructions for installation, use and maintenance to ensurethat the protection of the operator is not impaired. It is the responsibilityof the installer to ensure the safety and EMC compliance of any particularinstallation.2 SPECIFICATION2.1 MaterialsSample pipe work from flowmeter:- Stainless steelElectronic enclosure:- Stove enamelled steel for enclosed version.2.2 Ambient TemperatureSensor unit: 70°C MaxElectronics: 50°C Max.2.3 DisplayMulti-digit LCD - character height 12.7mm2.4 Display rangesDisplay range resolution: 0.1% from 10.0% to 100%; 0.01% from 0.01% to 9.99%2.5 Analogue outputs - isolated0 to 5 volts – minimum load 10k ohms. or4 to 20mA – maximum load 300 ohmsBottom scale is fixed at 0%, span is user programmable over the range of 5% to 100%2.6 Alarm outputsOne volt-free normally closed relay contact to indicate correct sensor andcatalyst temperature.2.7 AccuracyFor newly calibrated instruments ±2% of reading. Stability better than 2% of reading permonth. Repeatability better than 2% of reading.2.8 Speed of ResponseTypical T90 for combustion gas application is 5 seconds.2.9 Power Requirements110 or 240V ac 50/60 Hz at up to 200 VA.– voltage selector switch located on frontpanel of electronics unit.2.10 DimensionsSee diagrams