Ecom-D Page 17Re-connect the draft tubing. The display shows theexact measurement value which can be stored bypressing and added to those results pre-vious stored in the intermediate memory. The storedvalue is shown on the display. Press to exitthe draft measurement menu.8.3. Flow measurement (option)This measurement can be done with a pitot tube.First, the pitot factor of the pitot tube must be en-tered by going to / /). After connecting the pitot tube tothe analyzer, the zero point of the sensor can be setwith . Enter the cross section of the flow chan-nel with . This is needed for the calculation ofthe flow rate. After the pitot tube is positioned in theflow channel, the display shows the speed (m/s) andthe flow rate (Nm3/h). If the value is stable, press to store the value in the intermediatememory. If a printout of the values is needed duringthe recording to the intermediate memory, press. The complete content of the intermediatememory will be printed out.PressureDraught -0.12 hPaRecorded value:-0.12 hPaMeasurement stored inintermediate memoryMeasurement stored inintermediate memoryFlow measurementGas speed 0.3 m/sekM.Flow 44 Nm3/hConnections forpitot tube