Page 115. Power SupplyThe EN2-F is recommended to be powered by AC power whenavailable. The analyzer can also be powered using the internal bat-tery for a while (6 V; 3.8 Ah). The battery should be recharged whenthe analyzer starts beeping or an alert message appears. The bat-tery life can be checked by looking at the voltage indication on thedisplay in the Control menu. The low battery warning is activatedwhen the value “Batt” falls below 5.7 V. When the analyser is con-nected to AC power, the voltage will read as dashes “—.—“ Unplugthe power cord to read the battery voltage.When the battery voltage falls below 5.4 V, the instrument mustbe connected to AC power to retain power.6. Data Logging onto Memory CardThe multi-media card enables the storage of data records on a re-movable SD card. The data fields on the csv file are listed at the endof this manual. The following conditions must be fulfilled for using amulti-media card:- minimal card volume 32 MB - max. 2 GB- card formatted on 16 bit FATInsert Memory CardInsert the multi-media card as shown.Take care that the card inserts smoothlyand that you do not jam it in.Used batteries can be brought to recycling stations orpublic waste disposal companies!Never pull out cards during data recordingas this may cause a loss of data and/ordamage to the memory card itself.