User’s Manual Series TAS-1FEMA ELECTRÓNICA - Page 5Section 3.1 shows how to enter a code to select an inputsignal range. All codes are indicated on page 4.3- CONFIGURATIONExamples in this page show how to configure a TAS-1-IAC unitwith an input signal of 0/5 Aac and indication 0/2500.0. All otherunits and signal ranges follow the same procedures.Allows to link a current signal at the input terminalsto the High Indication Value memorized on the HIkeypad.1-Generate the high input level signal2-Press Adjust AdjThe instrument asks «Hi» or «Lo» ?3-Press Hi4-Message «Cr Hi» shows the unit has correctedthe high level value, linking the current input signalto the high indication value memorized (HI keypad)1-Generate the low input level signal2-Press Adjust AdjThe instrument asks «Hi» or «Lo» ?3-Press Lo4-Message «Cr Lo» shows the unit has corrected thelow level value, linking the current input signal value tothe low indication value memorized (LO keypad)3.2.5-CORRECTING THE LOW INPUT SIGNAL (OFFSEET)Allows offset correction, by linking the current input signal onterminals to the Low Indication Value memorized.VV1-Press Lo2- Current value for the low indication is displayed.Modify it by pressing on 1,2,3,4,5until we see 000003-Validate the changes pressing ↵↵↵↵↵message «INPUT» shows the configuration has beenaccepted3.2.2-ADJUSTING THE LOW INDICATION VALUEFor the low input level (0 Aac) we want toconfigure a low indication of 0000.0RVR3.2.1- DECIMAL POINT POSITIONActivate the Decimal Point at position XXXX.X1- Press DP2- message «DP» during 1 secondPress 2 to fix decimal point position 2(To fix at other positions, press 1,2,3,4 or 5)3- Validate changes pressinf ↵↵↵↵↵message «INPUT» shows the configuration hasbeen acceptedVRR3.2.4- ALARM SETPOINTSFix the setpoint for Alarm1 at 105001-Press AL2-Enter the number of the alarm we want tomodify 13-Current value for Alarm1 setpoint is displayed.Modify it by pressing on 1,2,3,4,5until we see 105004-Validate the changes pressing ↵↵↵↵↵message «INPUT» shows the configuration hasbeen acceptedVRRV3.2- DIRECT ACCESSThe frontal keypad hasseveral direct access buttons for aquick configuration of the instrument :Pad 1 (ADJ) - Correct the LOW or HIGH input levelsPad 2 (DP) - Decimal Point PositionPad 3 (LO) - Set for the LOW indication levelPad 4 (HI) - Set for the HIGH indication levelPad 5 (AL) - Alarms SetpointCONFIGURING A 0/5 AAC INPUT SIGNAL RANGE(Code 11 41 ; as shown on table on page 4, for aTAS-1-AAC unit)1- Press Progmessage «Prog» during 1 secondblank screen2- Input the code 11 413- Validate the code pressing ↵↵↵↵↵4- Message showing the selected range«5 A» during 1 secondVRVRThe TAS-1 units have several built-in predefined input signalranges, selectable with the help of codes (4 digit codes). Onpage 4 there is a list of all accessible ranges and the codesassociated.3.1- INPUT SIGNALVVV3.2.6-CORRECTING THE HIGH INPUT SIGNAL3.2.3- ADJUSTING THE HIGH INDICATION VALUEFor the high input level (5 Aac) we want to configurea high indication of 2500.01-Press Hi2- Current value for the high indication is displayed.Modify it by pressing on 1,2,3,4,5until we see 250003-Validate the changes pressing ↵↵↵↵↵message «INPUT» shows the configuration hasbeen acceptedVRR