5Signet PVDF Conductivity Sensors3/4 in. NPT orISO 7/1-R 3/4In aerated vessels install the electrode in a stillwell toprevent air from being trapped inside the electrode.Fill with3 to 4 in.of sealant1. Feed cable into watertight conduit.2. Apply thread sealant to the electrode before threading conduit onto electrode. Avoid twisting the cable.3. Secure cable with conduit or cable gland.4. For additional defense against possible accumulation of condensation at the back seal area of the electrode,fi ll the lower 75 mm to 100 mm (3 in. to 4 in.) of conduit or extension pipe with a fl exible sealant such as silicone.8850-3Conductivity Transmitterlocking ringconduit base8052 integraladapterelectrodeMax cable length15.2 cm (6 in.)• 3-8052 Integral Kit and 3-9000.392-X Liquid Tight Connector kit are required. (See Parts and Accessories on back page).• Cut the cable to approx. 15 cm (6 in.).• Strip outer cable cover back 5 cm (2 in.).• Strip each conductor to expose 1 cm (3/8 in.) of bare wire.• Tin each conductor with solder for best results.9900-1 Field Mountwith 8052 Integral Mounting Kitand 9900.396 Angle Adjustment AdapterIntegral InstallationSubmersible Installation