19APPENDIX B: Nebulizer Assembly & MaintenanceTo Disassemble Nebulizer: Remove Inlet (8) with hose from Housing (2). Remove Housing with theReservoir from the Manifold on the front panel with a pulling and twisting motion. Separate Reservoir (1)and Housing with a breaking and pulling motion. Remove the exposed aerosol generating parts bygrasping the Tab (3) with thumb and forefinger and pull Target Piece (4) straight out of the Housing.Remove Cap (5) and Dip Tube (7). Leave the Jet (6) and the O-ring (8) in place (DO NOT take the jet outfrom the nebulizer housing). See Figure 5-E & F p27.To clean or remove obstruction: Soap or detergent may be used to clean all of the above parts. Flushparts with water. Be careful not to enlarge any of the holes in the parts. Air-dry only before reassemble.To Assemble Nebulizer:Place BOTH of the o-rings over the Jet(6). Using dip Tube(7) as handle, insert the Cap(5) into the Jet(6)inside the Housing(2). Hold the Housing(2) with your left hand as shown by the above picture, whilebending the Dip Tube(7), push the Cap(5) with your right index finger into the underside of the O-ring untilit snaps. Grab the Tap(3) and slide the Target Piece(4) between the guides of Housing(2) and Cap(5) asfar as it will go. A few attempts may be needed for it to work. Snap Reservoir(1) back into Housing(2).Note:Always empty oil from the nebulizer and cover the opening with the provided orange cap (10) when it isnot being used to prevent dirt from getting in through the inlet opening.ORANGE COVER10HOUSING2RESERVOIR1TARGETPIECE4TAB3CAP5DIP TUBE7JET6INLET8Note: Nebulizer Assembly consists of part #3,4,5,6,7,9, 11only.O-RINGS9FILTER SCREEN11O-RINGO-RING