SpecificationsSpecifications are subject to change without notice.Table 3 General specificationsSpecification DetailsDimensions (W x D x H) 452 x 360 x 804 mm (17.8 x 14.2 x 31.7 in.)Enclosure Rating: NEMA 4x/IP56 (with closed doors)Material: PC/ABS case, PC door, PC hinges and latches, 316 SST hardwareIndoor use only. Keep away from direct sunlight.Weight 20 kg (45 lb) without reagents and standards, 36.3 kg (80 lb) with reagentsMounting Wall, panel or tableProtection class IPollution degree/installationcategory2/IIPower requirements AC: 100–240 VAC, 50/60 HzAC instrument: 0.5 A nominal, 8.3 A maximumAccessory output: 100–240 VAC, 5.0 A maximumDC: 24 VDCDC instrument: 1.2 A nominal, 9 A maximumAccessory output: 24 VDC, 1.6 A maximumConnection: 0.82 to 1.31 mm2 (18 to 16 AWG) wire, 0.82 mm2 (18 AWG)stranded recommended; field wiring insulation must be rated to 65°C (149 ºF)minimum, insulation rated for wet locations, 300 V minimum.Operating temperature 5 to 50 °C (41 to 122 °F)Operating humidity 5 to 95% non-condensingStorage temperature –20 to 60 °C (–4 to 140 °F)4–20 mA outputs Four; load impedance: 600 Ω maximumConnection: 0.33 to 1.31 mm2 (22 to 16 AWG) wire, 0.33 to 0.52 mm2 (22 to20 AWG) recommended, twisted pair shielded wireAlarm relay outputs Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VACmaximumConnection: 0.82 to 1.31 mm2 (18 to 16 AWG) wire, 0.82 mm2 (18 AWG)stranded recommended, insulation rated for 300 V minimum and for wetlocations to maintain enclosure ratings.Digital inputs Six; connection: 0.33 to 1.31 mm2 (22 to 16 AWG) wire, 0.33 to 0.52 mm2 (22 to20 AWG) stranded (isolated DC voltage input or an open-collector/relay contactclosure input) recommended, insulation rated for 300 V minimum and for wetlocations to maintain enclosure ratings.Fuses Input power—AC: T 1.6 A, 250 VAC; DC: T 6.3 A, 250 VACOutput power—AC: T 5.0 A, 250 VAC; DC: T 1.6 A, 250 VACAlarm relay outputs: T 5.0 A, 250 VFittings Sample line and sample bypass drain: 6 mm OD push-to-connect fitting forplastic tubingAir purge air inlet: 6 mm OD push-to-connect fitting for plastic tubingChemical and case drains: 11 mm (7/16 in.) ID slip-on fitting for soft plastic tubingEnglish 21