58 of 82 Operating the 9705 - 9705KSS - March 2006 - Edition 1Operator Manual HIAC9) Load a bottle of deionized water and flush the system by pressing the Flush icon toensure no sample remains in the inner tubing before running the next sample.4.5.1 Adding NotesFrom the Main Screen, the Notes tab is also available. To add notes about the sample, suchas information about the location, type, or any identifying information about the sample:1) Touch the upper-left hand corner of the main screen to invoke the Notes screen, shownin Figure 4-3.2) To add notes, touch the Add Notes icon to invoke a keyboard entry screen. Notes arelimited to 24 characters per line.3) Press Enter to return to the Notes screen.4) Press the Return icon to return to the main screen and start the run. The note willappear the next time a run is initiated.4.5.2 Editing Notes1) To edit a note, access the Notes screen by tapping the upper left corner of the mainscreen.2) Highlight the note and tap Edit. A keyboard entry screen appears. Notes are limited to24 characters.3) Press Enter to finish the entry and return to the Notes tab.4.6 Printing ResultsResults may be printed automatically or manually.4.6.1 Automatically Printing ResultsThe 9705 can automatically print results every time a sample run is completed.Fig 4-3 : Notes Tab