30TroubleshootingTable 7 Warning CodesDisplayedWarning Definition ResolutionEE SETUP ERR The EEPROM was corrupted. Values havebeen set to factory defaults. Contact the Service Department.EE RSRVD ERR The EEPROM was corrupted. Values havebeen set to factory defaults. Contact the Service Department.TEMP < 0 C The sensed temperature is below 0 °C (32 °F). Increase process temperature or discontinue use untilthe process temperature is above 0 °C (32 °F).TEMP > 50 C The sensed temperature is above 50°C(120 °F).Decrease process temperature or discontinue use untilthe process temperature is below 50 °C (120 °F).RED AMPL LOWSensor cap not installed or not installedcorrectly.Sensor not operating properly.Remove and reinstall sensor cap.Contact the Service Department.RED AMPL HIGHSensor cap not installed or not installedcorrectly.Sensor not operating properly.Remove and reinstall sensor cap.Contact the Service Department.BLUE AMPL LOWSensor cap not installed or not installedcorrectly.Sensor not operating properly.Remove and reinstall sensor cap.Contact the Service Department.BLUE AMPL HIGH Sensor cap not installed or not installedcorrectly. Sensor not operating properly.Remove and reinstall sensor cap.Contact the Service Department.