27Installation3.4.2 MountingWARNINGElectrocution hazard. If the cable and connector for the power supply are notaccessible after installation, an accessible local disconnection means for theinstrument power is mandatory.Figure 9 Panel mount bracket frame1. Cut an opening in the panel to accommodatethe bracket frame provided (this is the samesize as previous generations of ORBISPHEREtype 3600 instruments).Figure 10 Opening dimensions2. Install the provided frame in the opening3. Fold the 6 tabs over the panel lips, usingadjustable joint pliers.4. Slide the instrument in the bracket frame. Theinstrument should go over the four “T” pins.Rotate the 4 fast locking screws on both sides ofthe front panel and slide it in.5. Rotate the 4 fast locking screws 1/4 turn twice inthe lock direction as indicated on the side of thefront panel. This locks the instrument in place onthe four “T” pins.6. To access the connections inside theinstrument, remove the instrument housing (sixscrews on the back panel, and slide the housingback out)7. Pass the cables through the housing, thenthrough the cable gland (if applicable) and thenperform the connections as detailed below.Note: Do not forget to pass the cable through the housingbefore passing it through the cable gland on the backpanel.