16Operation4.4.2 SENSOR SETUPWIPE Initiates a wiping actionVERIFY Verification using CVMmodule Verification with dry standardCALIBRATESTANDARDSelection as perCAL. CONFIGOUTPUT MODECalibration with standardsolutionProcedure during calibration, menubasedOFFSETSelection as perCAL. CONFIGOUTPUT MODEZero point setting Zero point setting, menu basedCAL. FAKTORS FAC. STANDARD 0.50 to 2.00OFFSET -0.100 to +0.100 TRBFNUCAL. CONFIG OUTPUT MODEHOLDACTIVETRANSFERCHOICEBehaviour of the outputs duringcalibration or zero point settingCAL. INTERVAL Adjustable from 0 to 365 daysSET CAL DEFLT Reset to the default calibrationCONFIGUREEDITED NAME SET DEFAULTS Instrumentnumber Up to 16-character name possibleMEAS UNITS mg/L, FNU, NTU, TE/F,EBCSET DEFAULTSFNUCLEAN. INTERVAL SET DEFAULTS12 h10 min, 20 min, 30 min,2 h, 6 h, 12 h, 10:00RESPONSE TIME SET DEFAULTS15 s 0 to 60 sLOGGER INTERVAL SET DEFAULTS10 min 1-30 minRESOLUTION SET DEFAULTS0.xxx 0.xxx or 0.xxxx (<1 FNU)SET DEFAULTS Confirmation prompt Reset to default configuration for allmenu commands listed above.