4. Main parts RiseAtlasIn order to get access to the main parts, please see chapter 3.The main parts of RiseAtlas are:1. Hand controlThe RiseAtlas is turned Off/On with the hand control and also the liftmovement Up/Down is controlled by hand control.Please see chapter 9.1 Test and/or replace hand control.2. Electrical systema. Main Print Circuit Board (PCB),The PCB controls functionality of RiseAtlas.Please see chapter 9.3 Replace circuit boardb. Diagnostic circuit board,The diagnostic card controls statistics of RiseAtlas.Part of bottom plate and please see chapter 9.4 for replacement.c. Wire harness motorThe wire harness connects the different electrical components of RiseAtlas.Wire harness motor has component no.: 90002940d. Micro switch up,The micro switch up shuts off winding movement if H-adapter hits the switch or if the lift strap twist or tilts.Component no.: 90001167e. Micro switch down,The micro switch down shuts off winding movement when all lift strap is winded out and prevents the motor from wind-ing the lift strap the wrong direction. Component no.: 90001430f. Micro switch pre tension,RiseAtlas needs to be pre loaded in order to be maneuvered up or down and the switch secures that preloading isapplied and located behind PCB. Component no.: 90001430g. LED indicator,LED-indicators shows battery status and service need. Integrated part of bottom plate, please see chapter 9.4 forreplacement.9S E R V I C E M A N U A LSystemRoMedicTM