6. Troubleshooting6.2 Error messagesError Display Cause SolutionNo display None - No power connection- Blackout- Check the power connection- Check the main fuse inlaboratory and deviceNo runningUnknowRotor ID!- No rotor- Errors in drive or rotorrecognition function- Load a rotor- Restrat the device-Check the rotor sensor forforeign substance.HiTemperatureError!-The current temperatureis higher than the settemperature- Set the temperature higherLowTemperatureError!-The current temperatureis lower than the settemperature.- Set the temperature lowerThe lid is notopen None -Blackout - Emergency lid openThe lid is notclosedTo close thedoor- No working of lid pin- No power connection- Check the power connection- Close the lidOnacceleration,the device isvibrating** ImbalanceError! **-Rotor is loadedasymmetrically- Not tighten the rotorwith shaft- Load the rotor symmetrically- Tighten the rotor to themotor shaft.Failure ofpower supply** PowerFailure! ** - Failure of power supply - check the outletEtc** MotorOverheat! ** - Motor overheat - Stop the centrifugation- Let the device cool down** OverSpeed! **-The current speed ishigher than the set speed- Stop the centrifugation- Repeat the run15 l www.ihanil.com