44• Choose a homogeneous sample. Samples containing settleable solids need to be homogenized with ablender.• Preheat the Hanna Reactor HI 839800 to 150 °C (302 °F). For correct use of the reactor follow ReactorInstruction Manual.Use of the optional HI 740217 safety shield is strongly recommended.DO NOT USE AN OVEN OR MICROWAVE samples may leak and generate a corrosive and possibly explosiveatmosphere.• Remove the caps from two Reagent Vials.• Add exactly 2.0 mL of sample to one vial (sample vial), and 2.0 mL of deionized water to the othervial (blank vial), while keeping the vials at a 45-degree angle. Replace the cap tightly and mix byinverting each vial a couple of times.Warning: The vials will become very hot during mixing, be carefulwhen handling them.• Insert the vials into the reactor and heat them for 2 hours at 150 °C.• At the end of the digestion period switch off the reactor. Wait for twentyminutes to allow the vials to cool to about 120 °C.• Invert each vial several times while still warm, then place them in a testtube rack.Warning: The vials are still hot, be careful when handling them.• Leave the vials in the tube rack to cool to room temperature. Do not shakeor invert them, the samples may become turbid.BlankSampleBlankSampleBlankD.I. WaterSampleSampleCOD MR← white label → PreviousNext |