Icare ic100 tonometer manual ENGLISHwww.icaretonometer.com15GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITYIcare ic100 (TA011) is intended for use in the electromagnetic environment specified below.The customer or the user of the Icare ic100 (TA011) should assure that it is usuch an environment.Immunity test IEC 60601 Test level Compliance level Electromagnetic environment-GuidanceRadiated RF IEC 61000-4-3Conducted RFIEC 61000-4-63 V/m80MHz to 2,5 GHz3Vrms150 kHz to 80 MHz3V/m3VrmsPortable and mobile RF communications equipment should be used no closerto any part of the Icare ic100 (TA011), including cables, than the recommendedseparation distance calculated from the equation applicable to the frequencyof the transmitter.Recommended separation distanced = 1.2 √Pd = 1.2 √P 80 MHz to 800 MHzd = 2.3 √P 800 MHz to 2 5 GHzWhere P is the maximum output power rating of the transmitter in watts(W) according to the transmitter manufacturer and d is the recommendedseparation distance in metres (m).Field strengths from fixed RF transmitters, as determined by anelectromagnetic site survey should be less than the compliance levelin each frequency range.Interference may occur in the vicinity of equipmentmarked with the following symbol:NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radiobroadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, anelectromagnetic site survey should be considered. If the measured field strength in the location in which the Icare ic100 (TA011) is used exceeds the applicableRF compliance level above, theIcare ic100 (TA011) should be observed to verify normal operation. If abnormal performance is observed, additional measuresmay be necessary, such as re-orienting or relocating the Icare ic100 (TA011).Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND ICAREic100Icare ic100 (TA011) is intended for use in an electromagnetic environment in which radiated RF-disturbances are controlled. The customer or the user of theIcare ic100 (TA011) can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communicationsequipment (transmitters) and Icare ic100 (TA011) as recommended below, according to the maximum output power of the communications equipment.Rated maximum output power oftransmitter(W)Separation distance according to frequency of transmitter150 kHz to 80 MHzd= 1,2 √P80 MHz to 800 MHzd= 1,2 √P800 MHz to 2,5 GHzd= 1,2 √P0,01 0.12 0.12 0.230,1 0.38 0.38 0.731 1.2 1.2 2.310 3.8 3.8 7.2100 12 12 23For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using theequation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to thetransmitter manufacturer.NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.