572/573/KB/DS/FKB-BA-e-1774 409 TroubleshootingThe balance should be switched off for a short time following an interruption in the pro-gramme sequence and disconnected from the mains supply. It is then necessary to repeatthe weighing process from the beginning.Help:Interruption Possible causeWeight display is not illuminated. • The balance is not switched on.• The mains supply connection has been inter-rupted (mains cable not plugged in/faulty).• Power supply interrupted. .The weight display changes continu-ally• Draught/air movement• Table/floor vibrations• The weighing plate is in contact with foreignmatter.• Electromagnetic fields / static charging(choose different location/switch off interferingdevice if possible)The weighing result is obviously in-correct• The balance display is not set to zero• Adjustment is no longer correct.• Great fluctuations in temperature.• Electromagnetic fields / static charging(choose different location/switch off interferingdevice if possible)Switch the balance off if other error messages should appear and then switch on again.Contact the manufacturer if the error message does not disappear.10 Declaration of conformityTo view the current EC/EU Declaration of Conformity go to:www.kern-sohn.com/ceThe scope of delivery for verified weighing balances (= conformity-rated weighing balances) includes a Declaration of Conformity. Previous |