AEJ-C/AES-C/PLJ-C-BA-e-1514 4511.3 Weighing units switch-overThe weighing unit selected here remains even after disconnection from the mains.• Press UNIT repeatedly until the desired weighing unit is displayed.To enable or disable individual units go to the menu (See chap. 9.1 menu overview)For setting options see chap. 12.7Weighing scales with type approval options are restricted to [g] → [kg] →[ct].11.4 TaringThe dead weight of any weighing container may be tared away by pressing a button,so that the following weighing procedures show the net weight of the goods to beweighed. Put vessel of goods to be weighed on weighing plate and close the windscreen doors. Wait until the stability display appears ( ), then press TARE. The weight ofthe container is now internally saved. Weigh the goods to be weighed and close the wind screen doors. Wait until the stability display appears ( ). Read net weight.• When the balance is unloaded the saved taring value is displayed withnegative sign.• To delete the stored tare value, remove load from weighing pan andpress TARE.• The taring process can be repeated any number of times. The limit isreached when the whole weighing range is exhausted.11.4.1 Entering a tare value manually (pre-tare) In weighing mode press F button Press to select setting ENTER TARE and confirm by pressing PRINT. Use the arrow keys for entering the tare value and confirm by pressing PRINT. The entered tare value will be displayed with a “- “11.4.2 Deleting the tare valueTo delete the tare value, press the TARE-key or enter value“0.000“.