ALJ/ALS/PLJ/PLS-BA-e-1338327.2 Models with internal weight (KERN ALJ/PLJ)For non verified balances four adjustment possibilities are available in the menu.Menu settings In weighing mode press the MENU button and keep it pressed untilthe acoustic signal gets mute. displayed.Press MENUPRINT button.Use the MENU key to choose between the following settings:AUT-CAL Automatic adjustment with internal weight.Factory settings of models with type approval.I-CALAdjustment with internal weight after having pressedCAL; not available for models with settings appropriatefor verification.E-CALAdjustment with external weightnot available for models with settings appropriate forverification.TEC-CAL not documentedTake over selection using the PRINT button.The balance returns to menu.To finish the menu press the MENU button and keep it pressed untilthe acoustic signal gets mute. The balance returns automatically intoweighing mode.AUT-CALFactory setting ofverified balancesWith activated AUT-CAL function the internal adjustment isautomatically started when the balanceafter the weighing balance was disconnected fromthe mainsafter pressing ON/OFF in stand-by modeafter a temperature change of 1.5 °Cwith non loaded weighing plate / zero displayafter a time interval of 20 minwith non loaded weighing plate / zero displayThe automatic adjustment function is always enabled. You canstart adjustment at any time by pressing the CAL-key manually.