Zero point will be saved.SEL 15 kgThe balance will suggest aweighing value for adjustingweight. However, you may enter avalue of your choice via thenumber keys.Confirm weighing valueLQAD 15 kgCarefully place adjusting weight inthe centre of the weighing plateWait until the stability display shows:Adjustment process is started.The process of adjustment is completed.Remove adjusting weight, balance will returninto weighing mode automatically. In case ofan adjustment error or incorrect adjustingweight the display will show an errormessage (SPAN / FAKLD); repeat adjustmentprocess.SPAnPASSØ000000Ø999999Ø0.07.7.2 Models CPB.. M (verifiable models)For verified balances adjustment by means of a jumper is locked. In order to carryout adjustment the two contacts of the conductor plate must be shorted with a jumper(see chapt. 7.8.1).Observe stable environmental conditions. A warming up time of ca. 2 h is required forstabilization. Ensure that there are no objects on the weighing plate.CPB-BA-defsi-0712 51