26 CXB-BA-e-16228 Parts countingWith parts counting you can either count parts into a container or remove parts froma container. To count a greater number of parts the average weight per part has tobe determined with a small quantity (reference quantity). The larger the referencequantity, the higher the counting exactness. High reference must be selected forsmall parts or parts with considerably different sizes.8.1 Determination of the reference weight by weighingSet balance to zero and tare, as required.0�aPRE-TARE 0� 0��M+Place a known number of parts on the balance as reference weight118.3��aPRE-TARE 0� 0��M+If the "Weight" display is stable, enter the number of parts via number keys.The input is shown in the unit-weight window.118.3��aPRE-TAREWeight100�Piece weight1��M+QuantityWhile the “Number of pieces” display is flashing (3 sec) confirm by button.118.3��aPRE-TARE SAnnP� ------��M+Dead stop control is carried out and the calculated reference weight appears on thedisplay.118.3��aPRE-TAREWeight118.3��aPRE-TAREReference weight100��M+QuantityNow you can place the parts to be counted onto the weighing plate. All quantityparameters of your goods to be weighed are displayed: